ass 4 Pure Tone Audiometer

Audiolyser® ADL20
Portable, Reliable, Computerized |

The Audiolyser® ADL20 is a next-generation computerized digital audiometer designed for precise screening and monitoring of hearing function. Compact, portable, and reliable, it integrates advanced DSP technology with Audiowin®20 software to deliver accurate results, easy data management, and a user-friendly experience for healthcare professionals.

Audiolyser® ADL20

Audiolyser® ADL20 unique features

  • The Audiolyser® ADL20 is a computerized digital audiometer.
  • The electronics are integrated into the patient response button, rendering the device lightweight and easy to transport.
  • The Audiolyser® ADL20 can be set up with different headsets depending on user needs The DSP (Digital Signal Processor) located in the patient response button ensures communication with both the computer and the sound generation.
  • The Audiolyser® ADL20 is driven by the Audiowin®20 software, a simple and intuitive interface. Audiowin® 20 stores information in a data base, which can also be printed, recorded and exported to other software.
  • Storage of audiometric curves and results enables consultation of files at a later date as well
    as statistics processing of results.
  • With all its features, the Audiolyser® ADL20 is a reliable, high-performance and progressive
    tool.

Technical features

Audiolyser® ADL20 features

Modes of useManual or automatic
Sound transmissionContinuous, inversed or pulsed mode
Harmonic distortion± 3%
Frequency accuracy± 2.5%
Headset soundproofing10 – 40 dB (depending on model)
Length of lead3 metres
Storage temperature0 – 50°C
Temperature for use15 – 35°C
Humidity75% maximum
Altitude of operation< 2000 metres
Voltage5VDC (via USB port)
Supply current210mA maximum
Reference standardsEN 60601-1, EN 60601-1-2, IEC 60601-1-6, EN 62366-1, EN ISO 10993-1, EN ISO 10993-5, EN ISO 10993-10, NF EN ISO 14971, IEC 60645-1, ISO 8253-1, NF, ISO 389-1, ISO 389-8, ISO 7029, EN 62304, NF EN 20417, EN ISO 15223-1, NF EN ISO 13485
Type of audiometer4
Medical classIIa (rule 10)
GMDN Code Audiometer41187
Applied partBF Type
Dimensions in packaging255 × 210 × 100 mm
Total weight of device500 – 850 g approximately (depending on model)
Altitude of operation< 2000 metres
Year of 1st EC marking2010
NBUnder normal conditions of use, no heating time required for the equipment (§5.4 IEC 60645-1: 2012).

Specific features of each headset

TypeCalibration standardManufacturerHeadset weight
Sennheiser HDA 300ISO 389-5Manufacturer490 g
Holmco PD-81ISO 389-8Manufacturer725 g
RadioEar DD65ISO 389-8Manufacturer500 g

The ADL20® can output a maximum decibel (dB) threshold of 100 dB. The table below can be used to cross-reference the intensities in dB and the related frequencies in Hz.


Intensity limit for each headset (dB)

Frequencies (Hz)1252505007501000150020003000400060008000
dB max7080901001001001001001009080

Biocompatible single-use earpad covers

MaterialUnwoven PP (Polypropylene) 35g
Diameter11 cm
Reference standardsEN 62366-1, EN ISO 10993-1, EN ISO 10993-5, EN ISO 10993-10, NF EN ISO 14971, IEC 60645-1, ISO 8253-1, NF EN 1041+A1, EN ISO 15223-1
Medical classI (rule 1)
Earpad covers GMDN code63091