Computerised Spirometer

Spirolyser® Q13®
Easy, modern and quick |

The Spirolyser® Q13® is a computerised spirometer intended for use by health professionals. It uses single-use Fleisch type QFlow® sensors, ensuring accurate measurement of respiratory capacity while avoiding calibration, disinfection, and cross-contamination. The device enables Slow and Forced Vital Capacity, Maximum Voluntary Ventilation, and post-medication testing, making it a reliable tool for preventive respiratory medicine.

 

Spirolyser® Q13®

 

Spirolyser® Q13® features

TestsSlow and Forced Vital Capacity, Maximum Voluntary Ventilation, Post-medication
SensorFleisch type digital pneumotachograph
Flows Measurement Range-14 L/s to +14 L/s
Volume Measurement Range0 L to 10 L
Digital Resolution15 bits
Measurement Accuracy± 3% maximum
Lead Length3 metres
Storage Temperature0 – 50°C
Temperature for Use17 – 35°C
Atmospheric Pressure850 – 1060 hPa
Humidity75% maximum
Operational Altitude< 2000 metres
Voltage5VDC (via USB port)
Current200mW maximum
Reference StandardsEN 60601-1, EN 60601-1-2, IEC 60601-1-6, EN 62366-1, EN ISO 10993-1, EN ISO 10993-5, EN ISO 10993-10, NF EN ISO 14971, NF EN 62304/A1, ISO 20417, EN ISO 15223-1, NF EN ISO 13485, NF ISO 2859-1, ATS 2005
Medical ClassIIa
Software security classA
MarkingCE
Electrical classClass I
GMDN spirometer code35282
GMDN Qflow® code61097
Applied PartType BF (Qflow® sensor and device shell)
Size90x180x60mm
Device Weight250g
Year of first EC marking2015

QFlow® Sensor – product characteristics

Storage temperatureBetween 0 and 50°C
Operational temperatureBetween 17 and 35°C
Dimensions120x50x50 mm
Weight18 grams
Resistance at 14L/s1.35 cmH20/(L/sec)
Reference recommendationsATS 2005
Reference standardsEN ISO 10993-1, EN ISO 10993-5, EN ISO 10993-10, NF EN ISO 14971, ISO 20417, EN ISO 15223-1, NF EN ISO 13485, NF ISO 2859-1
Medical classIIa (rule 5)
CE MarkingCE
Applied partType BF (sensor)
Qflow® GMDN Code61097
EC marking acquired in2015

Nose clip characteristics

Storage temperatureBetween 0 and 50°C
Operational temperatureBetween 17 and 35°C
Reference standardsEN ISO 10993-1, EN ISO 10993-5, EN ISO 10993-10, NF EN ISO 14971, ISO 20417, EN ISO 15223-1, NF EN ISO 13485, NF ISO 2859-1
Medical classI (rule 1)
GMDN Code10907
Scroll to Top